Physical
Size: 20cm x 18.4cm x 4.8cm
Weight: 1.1 kg including Pad-Pak™ Battery
Defibrillator
Waveform: SCOPE® (self-compensating output pulse envelope) Biphasic waveform. Optimized biphasic waveform compensates energy, slope and envelope for patient impedance
Energy Selection
Pre-configured factory settings for escalating energy are AHA/ERC 2005
Adult 1. Shock 150J 2. Shock 150J 3. Shock 200J
Paediatric 1. Shock 50J 2. Shock 50J 3. Shock 50J
Patient Analysis System
Method: Evaluates patient’s ECG, signal quality, electrode contact integrity and patient impedance to determine if defibrillation is required
Sensitivity/Specificity: Meets AAMI DF80:2003 requirements
CPR Advisor
ICG technology incorporating ECG and impedance measurements.
Automatic CPR coaching/feedback to responder rate and depth of compressions with visual graph and audible voice prompts with a metronome.
Environmental
Operating Temperature: 0°C to 50°C (+32°F to +122°F)
Storage Temperature: -10°C to 50°C (+14°F to +122°F)
Relative Humidity: 5% to 95% (non-condensing)
Water Resistance: IEC 60529/EN 60529 IP56
Altitude: 0 to 15,000 feet (0 – 4,575 meters)
Shock: MIL STD 810F Method 516.5, Procedure I (40G’s)
Vibration: MIL STD 810F Method 514.5+
Category 4 Truck Transportation – US Highways
Category 7 Aircraft – Jet 737 & General Aviation (Exposure)
EMC: EN 60601-1-2, Second Edition: 2001
Radiated Emissions: CISPR11: 1997 +1A:1999+A2:2001 Group 1
Class B Electrostatic Discharge
Immunity: EN61000-4-2:1995 (8KV)+A1:1998+A2:2000
RF Immunity: EN61000-4-3:1996, 80 MHz – 2.5 GHz, (10V/m)+A1:1998 +A2:2000
Magnetic Field Immunity: EN61000-4-8:1993 (3 A/m)+A1:2000
Aircraft: RTCA / DO – 160D: 1997, Section 21 (Category M)
Falling height: 1 meter
Event Documentation
Type: Internal memory
Memory Capacity: 45 minutes of ECG (full disclosure) and event/incident recording Playback Capabilities: Custom USB cable directly connected to PC and Saver™ EVO Windows based data review software.
Materials Used
PAD SAM 500P: ABS, Santoprene. Printed circuit board with electronic components.
PAD Cartridge: Battery: Lithium Manganese Dioxide
Housing: ABS – Electrodes: Hydrogel, Silver, Aluminium and Polyester
Combined Electrode and Battery Cartridge
Pad-Pak™ Shelf Life: Check Expiration Date (typically 3.5 years from manufacture)
Weight: 0.44 lbs (0.2kg)
Size: 3.93 in x 5.24 in x .94 in (10cm x 13.3cm x 2.4cm)
Battery Type: Lithium Manganese Dioxide (LiMnO2) 18V, 0.8 Amp Hrs
Capacity: >30 shocks at 200J or 6 hours of continuous monitoring.
Adult Electrodes (Pad-Pak-01)
Standby Life: Typically 3.5 years from manufacture date
Electrodes: Samaritan® disposable defibrillation pads are
supplied as standard with each device
Placement: Anterior-lateral
Active Gel Area: 100cm2
Cable Length: 3.5 ft (1m)
Children Electrodes (Pad-Pak-02)
Standby Life: Typically 3.5 years from manufacture date
Electrodes: Samaritan® disposable defibrillation pads are supplied as standard with each device
Placement: Anterior-Posterior
Active Gel Area: 100cm2
Cable Length: 3.5 ft (1m) cable
Warranty
7 years
Maintenance
None required.
Self testing