Thermoline Refrigerated Incubator | TMLR-200 ...
The laboratory equipment plugs into an independent power outlet outside of the incubator. The temperature is factory set at 22°C, however the cabinet can operate between 20°C and 24°C in a 30°C ambient. The STAR 700 touch screen controller allows for accurate temperature control as well as a range of logging features such as alarms and temperature charts. The incubator has a key lockable glass door and password protected controller.
Australian Made Refrigerated Incubator
Safety Devices
The Incubator has a unique auto-test feature which allows for three main functions:
1) Alarm test: By accessing the temperature auto-test, the user is able to see a simulation of a high temperature alarm (24°C) and a low temperature alarm (20°C) without affecting the actual temperature inside the cabinet. This simulation will trigger the BMS relay so the remote monitoring of the alarms can also be checked.
2) Calibration test: This allows a Calibration Technician to check the linear accuracy over a 10.0°C range. Alarms will also be activated during this test.
3) Battery health test: Every 7 days there is an automated Battery Health Test. This test disconnects the battery charger and if the voltage drops below 22vDc in 5 minutes, an alarm is triggered notifying the operator that the battery has failed.
Specifications:
Temperature Range | 20°C to 24°C at ±1°C (factory set at 22°C) |
Temperature Control | Colour touch screen, Networkable, Control via PC if needed |
Alarms | Audible + visual. High, low, and power loss alarms |
Recording | Temperature + alarm recording (download via USB to CSV format) |
Internal Dimensions (WxDxH mm) | 600 x 550 x 530 |
External Dimensions (WxDxH mm) | 740 x 840 x 1100 |
Refrigeration | Internal equipment maximum heat output of 200 watts |
Electrical | 240V 4A |
Weight (kg when empty) | 90 |
To contact ENLAKE about Thermoline Refrigerated Incubator | TMLR-200 use Get a quote.
The laboratory equipment plugs into an independent power outlet outside of the incubator. The temperature is factory set at 22°C, however the cabinet can operate between 20°C and 24°C in a 30°C ambient. The STAR 700 touch screen controller allows for accurate temperature control as well as a range of logging features such as alarms and temperature charts. The incubator has a key lockable glass door and password protected controller.
Australian Made Refrigerated Incubator
Safety Devices
The Incubator has a unique auto-test feature which allows for three main functions:
1) Alarm test: By accessing the temperature auto-test, the user is able to see a simulation of a high temperature alarm (24°C) and a low temperature alarm (20°C) without affecting the actual temperature inside the cabinet. This simulation will trigger the BMS relay so the remote monitoring of the alarms can also be checked.
2) Calibration test: This allows a Calibration Technician to check the linear accuracy over a 10.0°C range. Alarms will also be activated during this test.
3) Battery health test: Every 7 days there is an automated Battery Health Test. This test disconnects the battery charger and if the voltage drops below 22vDc in 5 minutes, an alarm is triggered notifying the operator that the battery has failed.
Specifications:
Temperature Range | 20°C to 24°C at ±1°C (factory set at 22°C) |
Temperature Control | Colour touch screen, Networkable, Control via PC if needed |
Alarms | Audible + visual. High, low, and power loss alarms |
Recording | Temperature + alarm recording (download via USB to CSV format) |
Internal Dimensions (WxDxH mm) | 600 x 550 x 530 |
External Dimensions (WxDxH mm) | 740 x 840 x 1100 |
Refrigeration | Internal equipment maximum heat output of 200 watts |
Electrical | 240V 4A |
Weight (kg when empty) | 90 |
To contact ENLAKE about Thermoline Refrigerated Incubator | TMLR-200 use Get a quote.
Thanks for your help!
Sorry! Unable to submit your request at this time. Please try again later.
Report this product