Breas Airway Clearance Devices Nippy ...
TreatRepeat function allows clinicians to manually deliver different numbers of insufflations and exsufflations and then save the treatment they have just delivered to the device to be repeated automatically. Helping to improve accuracy, speed of set-up and titration
Width (cm) 29.7
Height (cm) 13.2
Length (cm) 22.3
MULTIPLE MODE OPTIONS
ADVANCED TITRATION TECHNOLOGY
ENHANCED PATIENT SYNCHRONY
Settings / Performance
Modes
Manual, Basic Auto, Program Mode, IPPB, NIV
Device Modes
Home, Clinical
Profiles
4 settable profiles
Insufflation Pressure
0 cmH2O to 70 cmH2O
Exsufflation Pressure
0 cmH2O to -70 cmH2O
Insufflation Time
0 to 5 seconds
Exsufflation Time
0 to 5 seconds
Pause Time
0 to 5 seconds
Flow Options (MI-E)
10% to 80% of Ti
Flow (IPPB)
5 L/min to 130 L/min
Inspiratory Trigger
Off or 1 to 9
Oscillations Frequency
Insufflations and or Exsufflations, Off – 4 – 20 Hz
Oscillations Amplitude
1 cmH2O to 10 cmH2O
Monitoring
Displayed data
Cough Peak Flow, Insufflation Volume, Pressure, Flow, SpO2, heart rate sessions used, average session duration
Waveforms
Pressure
Trends over 1, 6, 24 and 48 h via PC software
Power Supplies
Mains supply
100 to 240 V AC
External DC
Yes
Internal battery
Yes (optional)
Main alarms
Alarms
Circuit disconnection, running on battery power, external battery fault
Dimensions
W x H x D
285 x 280 x 200 mm
Weight
3.85 kg
Noise level (at 10 cmH2O constant pressure)
Less than 30 dB(A)
Connectivity and Software
Built-in connectivity
Ethernet, USB, SD card
Compatible Software
Prepared for Breas Cloud Service
Plug-and-Play accessories
SpO2 Hand held Remote Start/Stop and manual switch, manual foot pedal
Other
Circuits
Disposable leakage circuits with, disposable circuits with no leak
To contact Taleb Medical about Breas Airway Clearance Devices Nippy Clearway 2.0 use Get a quote.
TreatRepeat function allows clinicians to manually deliver different numbers of insufflations and exsufflations and then save the treatment they have just delivered to the device to be repeated automatically. Helping to improve accuracy, speed of set-up and titration
Width (cm) 29.7
Height (cm) 13.2
Length (cm) 22.3
MULTIPLE MODE OPTIONS
ADVANCED TITRATION TECHNOLOGY
ENHANCED PATIENT SYNCHRONY
Settings / Performance
Modes
Manual, Basic Auto, Program Mode, IPPB, NIV
Device Modes
Home, Clinical
Profiles
4 settable profiles
Insufflation Pressure
0 cmH2O to 70 cmH2O
Exsufflation Pressure
0 cmH2O to -70 cmH2O
Insufflation Time
0 to 5 seconds
Exsufflation Time
0 to 5 seconds
Pause Time
0 to 5 seconds
Flow Options (MI-E)
10% to 80% of Ti
Flow (IPPB)
5 L/min to 130 L/min
Inspiratory Trigger
Off or 1 to 9
Oscillations Frequency
Insufflations and or Exsufflations, Off – 4 – 20 Hz
Oscillations Amplitude
1 cmH2O to 10 cmH2O
Monitoring
Displayed data
Cough Peak Flow, Insufflation Volume, Pressure, Flow, SpO2, heart rate sessions used, average session duration
Waveforms
Pressure
Trends over 1, 6, 24 and 48 h via PC software
Power Supplies
Mains supply
100 to 240 V AC
External DC
Yes
Internal battery
Yes (optional)
Main alarms
Alarms
Circuit disconnection, running on battery power, external battery fault
Dimensions
W x H x D
285 x 280 x 200 mm
Weight
3.85 kg
Noise level (at 10 cmH2O constant pressure)
Less than 30 dB(A)
Connectivity and Software
Built-in connectivity
Ethernet, USB, SD card
Compatible Software
Prepared for Breas Cloud Service
Plug-and-Play accessories
SpO2 Hand held Remote Start/Stop and manual switch, manual foot pedal
Other
Circuits
Disposable leakage circuits with, disposable circuits with no leak
To contact Taleb Medical about Breas Airway Clearance Devices Nippy Clearway 2.0 use Get a quote.
Thanks for your help!
Sorry! Unable to submit your request at this time. Please try again later.
Report this product