Temperature | |
Working-temperature range |
5 °C above ambient temperature up to +50 °C Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors)
|
Setting temperature range | +18 to +50 °C |
Setting accuracy temperature | 0.1 °C |
Temperature |
2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
|
Humidity | |
Setting range active humidity control | 40 to 97 % rh and rh-Off |
Setting accuracy humidity | 0.5 % rh |
Humidity control |
Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 %
|
Control of standard components
|
|
CO2 control |
Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation
|
Adjustment range CO2 | 0 to 20 % CO2 |
Variation in time CO2 | +/- 0.2 % CO2 |
Adjustment range O2 | 1 to 20 % O2 |
Setting accuracy O2 | 0.1 % O2 |
Control technology
|
|
ControlCOCKPIT |
TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
|
Language setting |
German, English, Spanish, French, Polish, Czech, Hungarian
|
Function SetpointWAIT |
the process time does not start until the set temperature is reached
|
adjustable parameters |
temperature (Celsius or Fahrenheit), fan speed, CO2, programme time, time zones, summertime/wintertime
|
Communication | |
Interface | Ethernet LAN, USB |
Documentation |
programme stored in case of power failure
|
Programming |
AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
|
Safety | |
Autodiagnostic system |
integral fault diagnostics for temperature, CO2 and humidity limit control
|
Alarm | visual and acoustic |
Heating concept
|
|
6 sides |
large-area multi-function heating system on four sides with additional door and back heating to avoid condensation
|
Standard equipment
|
|
Works calibation certificate °C & CO2 |
incl. works calibration certificate (measuring point chamber centre) at +37°C, 5 % CO2 for standard units
|
Works calibration cerfificate °C, CO2 & rH |
incl. works calibration certificate for +37°C, 5 % CO2 and 90 % rh (requires option K7); standard equipment for units with active humidity control
|
Works calibration certificate °C, CO2, rH & O2 |
incl. works calibration certificate for +37°C, 5 % CO2, 90 % rh and 10 % O2 (requires option K7 and option T6); Standard equipment for units with O2 control
|
Internals |
1 pc stainless steel shelf, perforated
|
Internals |
1 pc stainless steel water dish
|
Door |
fully insulated stainless steel door with 2-point locking (compression door lock)
|
Door |
inner glass door with opening (8 mm Ø) to take gas sample
|
Stainless steel interior
|
|
Interior |
material 1.4301 (ASTM 304), corrosion resistant
|
Volume | 56 l |
Dimensions W x H x D in mm |
w(A) x h(B) x d(C): 400 x 425 x 330 mm
|
Max. number of grids/shelves | 5 |
Max. loading of chamber | 75 kg |
Max. loading per grid/shelf | 15 kg |
Textured stainless steel casing
|
|
Dimensions |
w(D) x h(E) x d(F): 559 x 791 x 521 mm
|
Housing | rear zinc-plated steel |
Electrical data | |
Voltage Electrical load | 230 V, 50/60 Hz approx. 1000 W |
Voltage Electrical load |
115 V, 50/60 Hz approx. 1000 W
|
Packing/shipping data
|
|
the appliances must be transported upright | |
Customs tariff number | 8419 8998 |
Country of origin | Federal Republic of Germany |
WEEE-Reg.-No. | DE 66812464 |
Dimensions approx incl. carton |
B x H x T: 730 x 950 x 640 mm
|
Net weight | approx. 55 kg |
Gross weight carton | approx. 74 kg |
To contact In Vitro Technologies about Memmert CO2 Incubator | 56L | MEMICO50 use Get a quote.
Product Specifications:
Temperature | |
Working-temperature range |
5 °C above ambient temperature up to +50 °C Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors)
|
Setting temperature range | +18 to +50 °C |
Setting accuracy temperature | 0.1 °C |
Temperature |
2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
|
Humidity | |
Setting range active humidity control | 40 to 97 % rh and rh-Off |
Setting accuracy humidity | 0.5 % rh |
Humidity control |
Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 %
|
Control of standard components
|
|
CO2 control |
Digital electronic CO2 control with dual beam NDIR system, with auto-diagnostic system and acoustic fault indication, barometric pressure compensation
|
Adjustment range CO2 | 0 to 20 % CO2 |
Variation in time CO2 | +/- 0.2 % CO2 |
Adjustment range O2 | 1 to 20 % O2 |
Setting accuracy O2 | 0.1 % O2 |
Control technology
|
|
ControlCOCKPIT |
TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
|
Language setting |
German, English, Spanish, French, Polish, Czech, Hungarian
|
Function SetpointWAIT |
the process time does not start until the set temperature is reached
|
adjustable parameters |
temperature (Celsius or Fahrenheit), fan speed, CO2, programme time, time zones, summertime/wintertime
|
Communication | |
Interface | Ethernet LAN, USB |
Documentation |
programme stored in case of power failure
|
Programming |
AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
|
Safety | |
Autodiagnostic system |
integral fault diagnostics for temperature, CO2 and humidity limit control
|
Alarm | visual and acoustic |
Heating concept
|
|
6 sides |
large-area multi-function heating system on four sides with additional door and back heating to avoid condensation
|
Standard equipment
|
|
Works calibation certificate °C & CO2 |
incl. works calibration certificate (measuring point chamber centre) at +37°C, 5 % CO2 for standard units
|
Works calibration cerfificate °C, CO2 & rH |
incl. works calibration certificate for +37°C, 5 % CO2 and 90 % rh (requires option K7); standard equipment for units with active humidity control
|
Works calibration certificate °C, CO2, rH & O2 |
incl. works calibration certificate for +37°C, 5 % CO2, 90 % rh and 10 % O2 (requires option K7 and option T6); Standard equipment for units with O2 control
|
Internals |
1 pc stainless steel shelf, perforated
|
Internals |
1 pc stainless steel water dish
|
Door |
fully insulated stainless steel door with 2-point locking (compression door lock)
|
Door |
inner glass door with opening (8 mm Ø) to take gas sample
|
Stainless steel interior
|
|
Interior |
material 1.4301 (ASTM 304), corrosion resistant
|
Volume | 56 l |
Dimensions W x H x D in mm |
w(A) x h(B) x d(C): 400 x 425 x 330 mm
|
Max. number of grids/shelves | 5 |
Max. loading of chamber | 75 kg |
Max. loading per grid/shelf | 15 kg |
Textured stainless steel casing
|
|
Dimensions |
w(D) x h(E) x d(F): 559 x 791 x 521 mm
|
Housing | rear zinc-plated steel |
Electrical data | |
Voltage Electrical load | 230 V, 50/60 Hz approx. 1000 W |
Voltage Electrical load |
115 V, 50/60 Hz approx. 1000 W
|
Packing/shipping data
|
|
the appliances must be transported upright | |
Customs tariff number | 8419 8998 |
Country of origin | Federal Republic of Germany |
WEEE-Reg.-No. | DE 66812464 |
Dimensions approx incl. carton |
B x H x T: 730 x 950 x 640 mm
|
Net weight | approx. 55 kg |
Gross weight carton | approx. 74 kg |