Large LCD display, the brightness can be adjusted to adapt to the clinical use of the environment
IPX4 high waterproof rating for all clinical application scenarios
Panel is easy to clean to ensure clinical cleanliness
Infusion speed can be adjusted at any time, even the pump is running
Large-scale LED colorful alarm light
Standard Package
IP300V Infusion Pump
A/C Adapter
User Manual
Technical Specification:
Fit IV Tubes
Can be used with any brand IV set after exactly calibration
Infusion Mode
Volumetric
Pumping Mechanism
Peristaltic finger pump
IV Sets
Standard IV sets of 10, 15, 20 and 60d/mL
Flow Rate Range
1.0 ~ 999.9 mL/h
Increment
0.1mL/h
Volume Limit
1 ~ 9999 mL
Flow Rate Accuracy
= ± 5% (after calibration)
Purge Rate
600 mL/h
KVO Rate
1 mL/h
Occlusion Pressure
Pressure detection device with 8 sensitivities setting, 40 ~ 140KPa.
Air-in-line Detection
Air-in-line detection device with 8 sensitivities setting, 50 ~ 1000µL.
IV Set Management & Flow Rate Calibration
Up to 10 adjustment values for various brands IV set could be stored to make the operation more convenient.
Alarms
Near empty, Syringe empty, Volume complete (when limit volume is set), Occlusion, Syringe loose, Operation error, Low Battery, Device failures, Three levels of alarm volume Low, Medium or High.
Large LCD display, the brightness can be adjusted to adapt to the clinical use of the environment
IPX4 high waterproof rating for all clinical application scenarios
Panel is easy to clean to ensure clinical cleanliness
Infusion speed can be adjusted at any time, even the pump is running
Large-scale LED colorful alarm light
Standard Package
IP300V Infusion Pump
A/C Adapter
User Manual
Technical Specification:
Fit IV Tubes
Can be used with any brand IV set after exactly calibration
Infusion Mode
Volumetric
Pumping Mechanism
Peristaltic finger pump
IV Sets
Standard IV sets of 10, 15, 20 and 60d/mL
Flow Rate Range
1.0 ~ 999.9 mL/h
Increment
0.1mL/h
Volume Limit
1 ~ 9999 mL
Flow Rate Accuracy
= ± 5% (after calibration)
Purge Rate
600 mL/h
KVO Rate
1 mL/h
Occlusion Pressure
Pressure detection device with 8 sensitivities setting, 40 ~ 140KPa.
Air-in-line Detection
Air-in-line detection device with 8 sensitivities setting, 50 ~ 1000µL.
IV Set Management & Flow Rate Calibration
Up to 10 adjustment values for various brands IV set could be stored to make the operation more convenient.
Alarms
Near empty, Syringe empty, Volume complete (when limit volume is set), Occlusion, Syringe loose, Operation error, Low Battery, Device failures, Three levels of alarm volume Low, Medium or High.