Working Mode
|
Rate/VTBI, Time/VTBI, Rate/Time, Micro, Simple Rate, Programmable, DOSE Weight, Drug Lab, AutoBolus, Manual Bolus.
|
Syringe
|
Open System for 5ml,10ml,20ml,30ml,50ml/60ml 5 kinds of syringe of different brand
|
syringe
management
|
could add / remove / edit / calibrate more than 50 kinds brand syringes
|
Accuracy
|
= ±2%,Built-in:= ±1%
|
Calibrate
|
Automatic or manual calibration of syringe accuracy
|
Rate
|
5ml syringe: 0.01-150 ml/h 10ml syringe: 0.01-300 ml/h 20ml syringe: 0.01-400 ml/h 30ml syringe: 0.01-600 ml/h 50ml (60 ml compatible) syringe: 0.01-1200ml/h
|
VTBI
|
0.1~9999 ml or OFF
|
Total Volume S
|
0-9999.9ml
|
KVO
|
0.1~20ml/h Adjustable,Default:1ml/h
|
Programmable
|
Maximum 8 steps of different syringe rate / time / VTBI
|
BOLUS
|
Manual Bolus:long Press the Bolus key to infuse and release to stop
Auto Bolus:short press the Bolus key and set the VTBI & Rate
|
Pressure
|
200mmHg - 1200mmHg, 20 level, live pressure display, mmHg & kPa two units
|
Pre-Alarm Occlusion
|
When the pressure in the tube goes up, it has the function of early warning
|
Back off
|
The pump will automatic release the pressure of syringe tube after occlusion in case the medical risk.
|
Pressure fast drop
|
Monitoring and controlling safety for the needle loose caused the pressure of syringe rapidly drop.
|
Pre-Alarm Finish
|
1 min~30 min, The alarm can be set ahead of VTBI
|
Fast start
|
First time install the syringe and when the pump start to working, it will automatic remove the
mechanical gap for high accuracy.
|
Parameters Modify
|
changes the Rate, Time, Dose, VTBI & other important parameters during the pump working
|
Keypad Locker
|
Manual / Auto lock keyboard
|
History Record
|
Record 20000 infuse records, 10000 alarm records, can be downloaded to the computer to save &query
|
Self-Test
|
safety live time self-test function to ensure the pump working properly
|
Alarm
|
Occlusion, Finish, Near Finish, Pause timeout, Pusher error, Lower Battery or damage, Motor error, Incorrect syringe type, handle clip error, parameters error and etc. 20 kinds safety monitoring alarm
|
Light Tips
|
3 color light tips:Green for pump working;Orange for low level alarm (counting working); Red for high level alarm (pump stop).
|
Voice Tips
|
Compliance YY 0709-2009 medical standard, Alarm and key volume level adjustable, with Mute function,
|
Monitor
|
High brightness color LCD, Adjustable brightness
|
Communication
|
RS485 port, connect with PC to update / download
|
AC Power
|
AC100V~240V,50Hz/60Hz,25VA
|
Battery
|
Ni-MH battery, DC9.6V, 2000mAh (after full charged, could continue work more than 7 hours by 5ml/h rate)
|
Battery power query
|
Live-time query battery life and battery charge status
|
Class
|
Class I, Type CF, IPX4
|
Installation
|
For round and square pole, Horizontal / Vertical, could superposed 1-3 units
|
Dimension
|
1.9Kg , 140mm×145mm×325mm
|
Environment
|
5?~40?; 20%~90% humidity; 70kPa~106kpa atmospheric
|
Use Range
|
For the patients with constant intravenous infusion of drugs
|